Chapter 7: Operation with the Filter Panel
9.
Configure the following for the stream, as needed. Press the appropriate button and enter the desired
value using the keypad that appears to the right, and then press Enter.
• Sample Flushes — Number of flushes to include
Pulses — Number of pulses to include
• Sample Flow Time — Delay time in seconds
10. Press the Sample Back-Flush button and then press either the Enabled or Disabled button that
appears. When enabled, this flush will occur at the end of every measurement.
11. Repeat this procedure for each stream installed with a filter panel.
GE Analytical Instruments ©2016
176 of 270
DLM 68100-09 EN Rev. A