Chapter 6: Analyzer Operation

Setting the Fail Safe Option

If the Fail Safe accessory is installed, the Analyzer is fitted with level sensors in each sample inlet to detect when
flow has stopped in any sample stream. Level sensors are installed in the reagent, dilution water, and Check
Standard carboys. These level sensors allow the Analyzer to automatically alert you when a carboy should be
refilled.
You must set the Fail Safe option to On in order to set the instrument to inform the Analyzer NOT to analyze a
stream that is not flowing. When flow resumes in that stream, the Analyzer will measure it again in its original
sequence. You can also set the Fail Safe option to Off to disable the Fail Safe option, which is helpful when
performing maintenance procedures.
To set the Fail Safe option
1.
Select the Maintenance tab.
2.
Press the Advanced button.
3.
Press the Fail Safe button, and then press either the On or Off button.

Setting the Keyboard Options

Specify the manner in which the Keyboard will be displayed: standard QWERTY mode or Alphabetical mode.
1.
Select the Maintenance tab.
2.
Press the Advanced button.
3.
The Keyboard button will either be set to Keyboard QWERTY or Keyboard Alphabetical. Press the
button to toggle between the two options.

Setting the Display Options

When the Analyzer values fall into the negative range, the user has the option to display these values as
Negative or as Non-Detect (ND).
1.
Select the Maintenance tab.
2.
Press the Advanced button.
3.
The Display Options button will either be set to Display Options Negative or Display Options ND.
Press the button to choose between the two options.