Chapter 5. Using the Configuration Menu

5.1.3 Calibrating Individual Channels

1. From the main screen, tap
Calibration screen.
2. In the Channel Calibration screen, enter the high and low values supplied with your dummy probes in their
respective boxes under External References. Tap the
values have been entered. Once this is completed, these values will be populated for every channel and do not
need to be re-entered for subsequent calibrations.
3. Select the Channel Number for the desired channel to configure. Make sure this is the channel to which your
calibration probes will be attached.
Lab Cal
4. Tap the
proceeding. The status should read "Connect probe to channel X, then click Read". Connect your low value
dummy probe to the M-Series input on the corresponding channel of the moisture.IQ module.
Read
5. Tap the
button. Wait up to a minute. When complete, the status will show "Reading complete on channel
X*."
6. Disconnect the low value dummy probe and connect the high value dummy probe. Do not change the external
reference values or tap the
Lab Cal
7. Tap
again, and wait for the message, "Connect the probe to channel X, then click Read." Tap
wait 2-3 minutes.
8. When complete, the new High and Low values will appear in their respective boxes under Channel Calibration.
You can now exit the Channel Calibration screen with the
channel requiring calibration.
62
Configuration,
Figure 56: Calibrating Channels
button to begin the low value calibration. The system will check for an active AutoCal before
Check
button.
then Probe Config, then Reference Values to access the Channel
Check
button on the Channel Calibration screen after both
Return
button, or repeat steps 3-8 for any other
Read
and
moisture.IQ User's Guide