I-6-8
Grid Non-Uniformity
Definition:
Serial Cmd:
Default:
Options:
I-6-9
Unused ECC
Definition:
Serial Cmd:
Default:
Options:
I-6-10 Grading Report
<VAL3>: Responds with a text report that summarized the grading of ISO/IEC15415.
This command will return the evaluation results, in a format similar to the one shown below.
V440-F C-Mount Code Reader User Manual
Grid Non-Uniformity is calculated in accordance with the ISO 15415
specification. The Grid Non-Uniformity grading is defined as follows.
A (4.0) if GN
038
B (3.0) if GN
0.50
C (2.0) if GN
0.63
D (1.0) if GN
0.75
F (0.0) if GN
>
0.75
When enabled, the Grid Non-Uniformity is attached to the symbol out-
put data as a value.
<K725,aperture,overall,contrast,modulation,reflectance margin,fixed
pattern damage,axial non-uniformity,grid non-uniformity,unused ECC>
Disabled
0 = Disabled
Unused ECC is calculated in accordance with the ISO 15415 specifica-
tion. The Unused ECC grading is defined as follows.
A (4.0) if UEC
062
B (3.0) if UEC
0.50
C (2.0) if UEC
0.37
D (1.0) if UEC
0.25
F (0.0) if UEC
<
0.25
When enabled, the Unused ECC is attached to the symbol output data
as a value.
<K725,aperture,overall,contrast,modulation,reflectance margin,fixed
pattern damage,axial non-uniformity,grid non-uniformity,unused ECC>
Disabled
0 = Disabled
1 = Enabled
1 = Enabled