Storing the hardware

Return the angular position measurement kit components to the kit case
when you have finished using them.
Before you store the precision index table:
1. Install the locking ring and bolt to hold the top and bottom together.
Tighten the bolt securely.
2. After performing step 1, disengage the locking loops. This helps extend
their useful life.

Maintenance

Precision index table

The precision index table's top and base have 360 matching teeth
comprising a Hirth coupling. For best performance of the table, these
teeth should be as clean as possible. You should visually inspect the teeth
each time you use the table. The procedure below is recommended for
cleaning Hirth coupling teeth.
BE SURE TO WEAR APPROPRIATE PROTECTIVE SHIELDING
AND CLOTHING WHEN WORKING WITH COMPRESSED AIR
AND ACETONE.
With the precision index table's top and bottom separated, use compressed
air to blow loose material from the Hirth coupling teeth.
If necessary, use an acid brush to remove material from the teeth.
If necessary, use acetone and the acid brush to remove material from the
teeth.
Measurements Reference Guide
4-67