Chapter 4 Angular Measurements
Mounting and Aligning the Optics on the Target Machine
Assemble the mounting hardware and optic you plan to mount on the
table. Follow these steps:
a. Assemble the base, post, and height adjuster (Figure 4-5).
b. Attach either the interferometer or the reflector to this assembly.
If you plan to mount the interferometer on the table, attach the
interferometer to this assembly as shown in Figure 4-5 (pitch
measurements) or Figure 4-6 (yaw measurements).
If you plan to mount the reflector on the table, attach the reflector
to this assembly as shown in Figure 4-7 (pitch measurements) or
Figure 4-8 (yaw measurements).
Mount the optic you assembled in step 5 on the machine's table top.
Keep the following in mind (see Figures 4-3 and 4-4):
The side of the interferometer with the two openings must face the
reflector.
One of the single openings on the interferometer must face the laser
head.
If you plan on making a straightness or squareness measurement later,
position the height adjuster flush with the top of the post so you can easily
change optics.
4-10
Measurements Reference Guide