Appendix A Operating Specifications
Introduction
This appendix lists the operating specifications for the Agilent 5530
Dynamic Calibrator.
Table A-1. General specifications
Characteristic
Axis velocity, using Linear
optics (Agilent 10766A
Interferometer).
Power requirements—
electronics (via USB)
Data rate
A-quad-B differential input
threshold
A-quad-B differential input
impedance
A-quad-B input rate
A-2
Specification
700 mm (27.6 in) per second, using Agilent 5519A Laser
Head
1000 mm (39.4 in) per second, using Agilent 5519B Laser
Head
(500 mA USB 2.0 hub required)
10 kHz
±0.5 V minimum;
±7.0 V maximum
100Ω
> 2 ns edge-to-edge, or < 500 kHz information rate
Example: If A-quad-B is moving at maximum speed, A and B
both must be < 125 kHz.