Appendix A Operating Specifications
Table A-3. Environmental compensation specifications
Characteristic
Maximum compensation
update rate
Wavelength of light (WOL)
2
compensation
Air temperature operating
range
Absolute pressure
Agilent E1738A air sensor
heat dissipation
Agilent E1738A air sensor
time constants
Material Temperature
Compensation
Material temperature
sensors operating range
Agilent E1737A material
temperature sensor time
constant
A-4
Specification
1
per 15s
Automatic calculation or Manual entry.
Automatic calculation requires Agilent
E1738A Air Sensor. Air pressure,
temperature, relative humidity, and computed
WOL provided on the display. If multiple
sensors are used, average values are
displayed.
Manual entry is via keyboard.
Range: 0.1000000 to 1.0000000 ppm.
0° to 40° C (32° to 104° F)
70 to 110 kPa (10–16 psia)
20 mW maximum
2 mW typical
5 minutes typical for air temperature; less
than 1 second typical for absolute pressure
Automatic calculation or Manual entry.
Automatic calculation requires 1 to 3
Agilent E1737A Material Temperature
Sensors. Output of each connected sensor, and
average of all, is provided on the display.
Manual entry is via keyboard.
0° to 40° C (32° to 104° F)
3
40 seconds typical
Measurements Reference Guide