Agilent Technologies 5530 Reference Manual page 142

Dynamic calibrator
Hide thumbs Also See for 5530:
Table of Contents

Advertisement

Chapter 4 Angular Measurements
Angular Position Measurements
Step
Art
1
Remove optic, if necessary.
2
1
Remove four M4 cap screws holding retainer ring to fixture top plate.
3
1
Remove clutch control shaft from retainer ring.
4
2
Remove clutch washer from base of clutch shaft.
5
3
Cut clutch washer and remove from shaft.
6
4
Slide new clutch washer onto shaft—paper-covered adhesive side first.
7
Remove protective paper from adhesive.
8
Press washer onto base.
9
1
Replace shaft in retainer ring.
10
1
Fasten retainer ring to fixture top plate.
Figure 4-34. Replacing the clutch washer
4-70
Procedure
1
2
3
Measurements Reference Guide