NOTE
Chapter 4 Angular Measurements
Making the Measurement
Figure 4-14. Collect Data: ANGULAR screen
Move the machine's moveable part to the first measurement point.
Select Record.
If you are using the optional remote control unit, press the Record button.
Continue moving the machine's moveable part to each point and recording
measurements until the machine has reached the last measurement
position.
If at any point the Reset Laser message flashes, the laser beam path has
been broken. Move the moveable optic back to the last point measured,
then select Reset Position to reset the system.