Agilent Technologies 5530 Reference Manual page 98

Dynamic calibrator
Hide thumbs Also See for 5530:
Table of Contents

Advertisement

1a
1b
Y
Z
1
1
Table "neutral" position
1a
No force on clutch
For the machine control
program shown in Figure 4-18,
the clutch is depressed 1 mm
from this position. This motion
ensures appropriate
engagement force between the
table's top and base.
1b
Table top engaged with base
Chapter 4 Angular Measurements
Angular Position Measurements
2a
2b
X
2
Table "disengaged" position
2a
Lift top 3 mm to allow base to
rotate. First mm lift returns to
"neutral" position and re-
engages the clutch; remaining
2 mm provides clearance
between table's top and base,
to allow base to rotate. Set the
machine's reference point
(relative zero) for this axis after
you have performed this lift.
2b
Top lifted to allow base to
rotate, as described above.
Figure 4-16A. Angular position measurements—principles (Part 1 of 2)
4-26
3a
3b
Y
Z
X
2
3
Table base rotated
3a
Table top has been disengaged, as
described in 2a.
3b
Machine part being calibrated is
rotated, turning the table's base with it.
Measurements Reference Guide
Y
Z
X
3

Advertisement