Chapter 9 Flatness Measurements
Completing the Set Up Flatness Screen
Use the procedures in Chapter 2 regardless of the length of your
measurement line.
Instead of moving the machine's moveable part, move the reflector.
You are now ready to complete the Set Up FLATNESS screen.
Completing the Set Up Flatness Screen
Before checking the alignment for beam strength, you must complete the
Set Up FLATNESS screen. This screen allows you to supply the software
with general measurement setup information, such as the foot spacer
length. To complete this screen, follow these steps:
1
On the Other Measurements screen, select Flatness.
The Set Up FLATNESS screen is displayed (Figure 9-12).
2
In the Measurement Units box, select either English or Metric.
Your selection indicates the units for angles and for foot spacer and line
lengths. It also determines the display units for the diagonal length on the
Set Up FLATNESS screen and the Laser Error and Target position units
on the Collect Data: FLATNESS screen.
9-26