AGFA Drystar 5300 Reference Manual page 117

Hide thumbs Also See for Drystar 5300:
Table of Contents

Advertisement

2
To determine the reference values for geometry, measure the dimensions A
and B of the geometry square on the test image.
Make sure to measure distance A from the left edge of the left line to the
right edge of the right line and distance B from the upper edge of the
upper line to the lower edge of the lower line.
We strongly recommend using a 30-cm (12-inch) machinist scale with
0.5 mm divisions (1/64 inch).
3
Record these values as reference dimensions A
Drystar 5300 Chart 4 ('Drystar 5300 Geometric Consistency Control Chart').
Refer to Appendix C
These charts will be used for the annual quality test. For more information, refer to
'Performing the annual QC test (Geometric Consistency Test)'
4
Save this film for future reference.
2920B EN 20031206
'Quality Control Charts'
and B
ref
on page 199.
on page 121.
on the
ref
117

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents