AGFA Drystar 5300 Reference Manual page 90

Hide thumbs Also See for Drystar 5300:
Table of Contents

Advertisement

$ If you select "CubicBell", the following screen will appear, allowing you to
adjust the smoothing factor:
Enter smoothing
parameter:
advised value = 0.35
range [0.1-10]
enter absolute value
y.yy
Enter the smoothing factor value by means of the arrow keys. Refer to
on page 27. Press the Confirm key to store the data, or the Escape key to quit the
procedure without any changes.
Changing the maximum density (MaxD)
1
Perform steps 1 to 3 of the
procedure on page
The 'Select for Modality' screen is displayed:
Select for
Modality16char:
1 Look-up table
2 Interpolation
3 MaxD
2
In the 'Select for Modality' screen, press the Down key twice, followed by the
Confirm key to select 'MaxD'.
The 'MaxD (maximum density)' screen is displayed:
Select MaxD for
Modality16char:
2.80
2.85
...
* 3.10
3.15
...
3.40
All maximum densities are selectable starting from 2.80 to 3.40 in steps of 0.05
units. The activated setting is marked with a '*'.
" The maximum density will be limited to the maximum density of the media.
3
Press the up/down key to select the desired maximum density setting for this
modality.
4
Press the Confirm key to store the data, or the Escape key to quit the
procedure without any changes.
90
CS
'Changing Image quality settings for a modality'
87
to select the modality and choose 'Select other'.
CS
CS
'Data entry'
2920B EN 20031206

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents