AGFA Drystar 5300 Reference Manual page 87

Hide thumbs Also See for Drystar 5300:
Table of Contents

Advertisement

Changing Image quality settings for a modality
1
Perform steps 1 and 2 of the
procedure on page 59.
2
On the 'Change settings' screen, press the Down key three times, followed by
the Confirm key to select 'Image quality'.
The following screen will appear:
1 General setting
2 Def. densitometer
3 Modality16char
4 Modality16char
5 etc...
3
Press the Up/Down key, followed by the Confirm key to select the modality for
which you want to change the Image quality settings.
" If the settings for the selected modality were set via a remote computer,
you will not be able to change the image quality information for this
modality manually. In this case the following screen will appear:
Modality16char
Parameters can
only be changed by
DICOM configurator
If the image quality parameters need to be changed for this modality, please use the
DICOM Configuration tool, in the specialist mode on the remote PC. Alternatively,
you can remove this modality (refer to
new modality with the same name. For this 'new' modality new parameters can be
assigned via the local keypad (refer to
" If a file with default settings ('Profile default') for the selected modality
exists, the following screen is displayed:
Default for
Modality16char:
1 Profile default
2 Select other
" If you want to use these profile default settings for the current modality,
press the Confirm key. The Drystar 5300 will be loaded with these
settings and return to the 'Image quality' main menu.
2920B EN 20031206
CS
CS
CS
'Accessing the Change settings menu'
'Removing a modality'
'Adding a modality'
on page 79) and add a
on page 80).
87

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents