Establishing the daily operating reference density levels
This procedure enables you to establish the base line values for:
Base+Fog density,
Mid density,
DD (Density Difference),
MaxD (maximum density).
By default, the internal densitometer of the Drystar 5300 will be used. The
internal densitometer of the Drystar 5300 is calibrated at installation.
Authorized
densitometer annually or after major service or repair.
Do not change the 'default' internal densitometer during the entire QC
procedure!
To establish the daily operating density levels, proceed as follows:
1
Press the Key-operator key to enter the key-operator mode.
2
Press the Down key seven times, followed by the Confirm key to select
'Quality control'.
1 Show settings
2 Change settings
3 Print image
4 Save configuration
5 Restore config.
6 Calibration
7 Service actions
8 Quality control
9 Installation
A confirmation screen appears:
Start QC
Procedure
YES (v)
NO (x)
3
Press the Confirm key (YES) to start the QC procedure or the Escape key
(NO) to quit.
QC procedure
in progress
Please wait
The Drystar 5300 will automatically print the QC test image.
2920B EN 20031206
service
personnel
KO
QC
QC
62%
should
recalibrate
the
internal
111