AGFA Drystar 5300 Reference Manual page 105

Hide thumbs Also See for Drystar 5300:
Table of Contents

Advertisement

4
When the film calibration has been completed successfully, three cases are
possible:
Calibration successful, results OK.
The obtained MaxD (3.10) after calibration, is displayed:
Film
calibration OK
MaxD = 3.10 (xxxxx)
(xxxxx) is the code of the Densitometer type (refer to
on page 86).
Calibration successful, but MaxD differs more than 2% from the target value.
The obtained MaxD (3.10) after calibration, is displayed:
Retry
calibration or
Resume printing
MaxD = 3.10 (xxxxx)
(xxxxx) is the code of the Densitometer type (refer to
on page 86).
Calibration successful, but the maximum density has been reached.
The obtained MaxD (3.10) after calibration, is displayed:
Film
calibration
MaxD is reached
MaxD = 3.10 (xxxxx)
(xxxxx) is the code of the Densitometer type (refer to
on page 86).
5
Press the Confirm key to return to the Calibration menu.
When the calibration process is interrupted, you will have to start the
calibration procedure again. Otherwise the old calibration parameters
will be used.
2920B EN 20031206
Flashing
CA
Flashing
Flashing
CA
Flashing
Flashing
CA
Flashing
'Changing the default densitometer'
'Changing the default densitometer'
'Changing the default densitometer'
105