Always take into account the markings provided on the inside and outside of
the printer. A brief overview of these markings and their meaning is given
below.
Safety warning, indicating that the Drystar 5300 manuals should be
consulted before making any connections to other equipment. The
use of accessory equipment not complying with the equivalent
safety requirements of this printer may lead to a reduced level of
safety of the resulting system. Consideration relating to the choice of
accessory equipment shall include:
• Use of the accessory equipment in the patient vicinity,
• Evidence that the safety certification of the accessory equipment
In addition all configurations must comply with the medical electrical
systems standard IEC 601-1-2. The party that makes the
connections acts as system configurator and is responsible for
complying with the systems standard.
If required contact your local service organization.
Caution hot:
Keep hands clear from the thermal print head.
In order to reduce the risk of electric shock, do not remove any
covers.
Type B equipment:
Indicates that the Drystar 5300 complies with the limits for type B
equipment.
Supplementary protective earth connector:
Provides a connection between the Drystar 5300 and the potential
equalization busbar of the electrical system as found in medical
environments. This plug should never be unplugged before the
power is turned off and the power plug has been removed.
Intergrounding connector:
Provides a connection between the printer and other equipment
which might exhibit minor ground potential differences. These
differences may degrade the quality of communication between
different equipment. Never remove connections to this terminal.
Protective earth (ground):
Provides a connection between the printer and the protective earth
of the mains. Do not remove this connection, because this will have
a negative influence on the leakage current.
2920B EN 20031206
has been performed in accordance with the appropriate IEC 601-1
and IEC 601-1-2 harmonized national standard.