AGFA Drystar 5300 Reference Manual page 89

Hide thumbs Also See for Drystar 5300:
Table of Contents

Advertisement

Changing the interpolation
1
Perform steps 1 to 3 of the
procedure on page
The 'Select for Modality' screen is displayed:
Select for
Modality16char:
1 Look-up table
2 Interpolation
3 MaxD
2
In the 'Select for Modality' screen, press the Down key, followed by the
Confirm key to select 'Interpolation'.
The 'Interpolation' screen is displayed:
Select for
Modality16char:
1 Replication
2 Bilinear
3 None
4 Cubic Hi-Res
5 CubicBell
3
Press the up/down key to select the desired interpolation setting for this
modality.
$ If you select "Cubic Hi-Res" the following screen will appear, allowing you to
adjust the smoothing factor:
Enter smoothing
parameter:
advised value = 2.50
range [-5.00-0.00]
enter absolute value
y.yy
Enter the smoothing factor value by means of the arrow keys. Refer to
on page 27. Press the Confirm key to store the data, or the Escape key to quit the
procedure without any changes.
2920B EN 20031206
'Changing Image quality settings for a modality'
87
to select the modality and choose 'Select other'.
CS
CS
CS
'Data entry'
89