Film calibration
The film calibration combines MaxD and sensitometry calibration and is
activated as one single procedure. Separate MaxD and sensitometry
calibration steps are not required.
1
Perform steps 1 to 2 of the
page
103
2
On the Print image menu, press the Confirm key to select 'Film calibration'.
The 'Film calibration' screen is displayed:
Film calibration
in progress
Please wait
The counter indicates the progress of the calibration process. The calibration film is
measured automatically by the built-in MDM (Macro Densitometer). The
Drystar 5300 is being calibrated during the printing process.
" If the Drystar 5300 cannot read the Film identification tag, a calibration is
impossible and the following message is displayed:
Unidentified media
in input tray
No calibration
allowed.
Press the Confirm key to retry calibrating.
3
If the film calibration failed, the following message will appear.
Film calibration
failed
Please retry
Press the Confirm key to retry return to the calibration menu.
104
to select the 'Calibration' screen.
CA
62%
CA
Flashing
CA
Flashing
'Accessing the Calibration menu'
procedure on
2920B EN 20031206