12.USE OF EQUIPMENT IN WORK AREAS WHERE EXPLOSIVE ATMOSPHERES MAY
OCCUR. CE Directive 94/9/EC – ATEX Directive – July 2003
(a) LOCATION OF EQUIPMENT. This equipment is certified as category 3D for use in
zones designated 22 only.
(b) INSTALLATION AND MAINTENANCE. Electrical enclosures should be inspected and
maintained on a regular basis to prevent the build up of dust deposits and to ensure
the integrity of seals.
Before undertaking maintenance the equipment must be disconnected from the
source of electrical power.
During installation and maintenance care should betaken to prevent the ingress of
dust into any electrical enclosure. Should dust ingress occur then all deposits must be
removed before the equipment is put into service.
Following installation and maintenance and before putting into service all electrical
enclosures and cable glands are to be correctly secured to prevent dust ingress.
Care must be taken during installation and maintenance not to damage door seals
and cable entry glands.
(c) USE. To ensure safe operation deposits of dust on any surface should be removed
immediately to prevent an increase in the surface temperature of equipment.
13.HEIGHT ADJUSTMENT. Care should be taken when adjusting the height of the Pipeline
and Pharmaceutical detectors. Lower the height adjustment to its minimum before
moving. If appropriate measures are not taken, severe injury to an individual or damage to
the detector could result.
14.LIABILITY. This machinery contains high voltages of a hazardous and potentially fatal
nature. Spectrum Inspection Systems Limited cannot accept any liability for death or
bodily injury resulting from improper work undertaken by unqualified operatives, or due to
deviations from the maintenance instructions within this manual.
This Liability statement is in addition to the terms of sale.