Page 1
BATCH LENGTH MENUS A to Z L o m a S u p e r s c a n M i c r o I S C *BTCH LENTH 1.00 RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT SETUP RESULTS CALIBRATE CHANGE RUN ERROR Description:...
Page 2
BATCH START MENUS A to Z L o m a S u p e r s c a n M i c r o I S C *BTCH START 1.30 RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT RESULTS SETUP CALIBRATE CHANGE RUN ERROR Description:...
Page 3
BATCH UNITS MENUS A to Z L o m a S u p e r s c a n M i c r o I S C *BTCH UNITS PACKS RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT SETUP RESULTS CALIBRATE CHANGE RUN ERROR Description:...
Page 4
CALENDAR MENUS A to Z L o m a S u p e r s c a n M i c r o I S C CALENDAR RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT RESULTS SETUP CALIBRATE CHANGE RUN ERROR Description: The date and time specified within this menu are used to time stamp the data in the contaminant, fault and error logs, and also on the printed reports.
Page 5
Report Options menu. SAMPLE REPORT --------------------------------------- : LOMA SUPERSCAN MICRO METAL DETECTOR : --------------------------------------- CALIBRATION REPORT Time : 15-32-16 Date : WED 20 SEP 1995...
Page 6
CHANGE PRODUCT MENUS A to Z L o m a S u p e r s c a n M i c r o I S C ANY DISPLAY RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT RESULTS SETUP CALIBRATE CHANGE RUN ERROR Description: In order to change the running product press the [CHANGE] key, then use the [ ] or [ ] to make...
Page 7
CLEAR CONTAMINANTS LOG MENUS A to Z L o m a S u p e r s c a n M i c r o I S C CLEAR CONTAMS? RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT SETUP RESULTS CALIBRATE CHANGE RUN ERROR Description:...
Page 8
CLEAR ERRORS MENUS A to Z L o m a S u p e r s c a n M i c r o I S C CLEAR ERRORS? RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT RESULTS SETUP CALIBRATE CHANGE RUN ERROR Description: Clearing a run error will turn the run error LED and Auxiliary relay off if it is configured for run error...
Page 9
Clearing the faults as described here will reset the LED fault indicator and the internal fault relay, but the details in the fault log will not be cleared. These are maintained by the system for Loma Customer Service Engineer reference.
Page 10
This value is automatically derived during a calibration cycle, and should therefore not normally require adjustment: only alter the value if advised to do so by a LOMA Service Engineer. Press the [ENTER] key to view the compensation and signal level values; the [ ] and [ ] keys to adjust the compensation value.
Page 11
If the detector suffers from memory corruption it will reset itself automatically into the top of this menu showing the message +NOT DETECTING+. Under these circumstances please consult your nearest LOMA Service Centre. If BATCH mode operation for PV Test is selected then the Configuration menu cannot be entered...
Page 12
CONTAMINANTS MENUS A to Z L o m a S u p e r s c a n M i c r o I S C CONTAMINANTS RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT RESULTS SETUP CALIBRATE CHANGE RUN ERROR Description: From within this menu the contaminant log can be viewed, cleared and, if the reports option is enabled, a contaminant report sent down the serial link.
Page 13
To gain access to this command, the report must first be enabled in the Report Options menu. SAMPLE REPORT --------------------------------------- : LOMA SUPERSCAN MICRO METAL DETECTOR : --------------------------------------- METAL CONTAMINANTS REPORT Time : 15-39-25 Date : WED 20 SEP 1995...
Page 14
DELAY MENUS A to Z L o m a S u p e r s c a n M i c r o I S C *DELAY 1.25S RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT RESULTS SETUP CALIBRATE CHANGE RUN ERROR Description: The delay is the time taken for a contaminated product to travel from the detector head to the reject point, or from the photo eye (PEC) if one is fitted to the system.
Page 15
DISABLE REJECT MENUS A to Z L o m a S u p e r s c a n M i c r o I S C DISABLE REJECT? RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT SETUP RESULTS CALIBRATE CHANGE RUN ERROR Description: Pressing the [ENTER] key will cause the reject device to be “turned off”...
Page 16
DRY PRODUCT MENUS A to Z L o m a S u p e r s c a n M i c r o I S C *DRY PRODUCT YES RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT RESULTS SETUP CALIBRATE CHANGE RUN ERROR Description:...
Page 17
DWELL MENUS A to Z L o m a S u p e r s c a n M i c r o I S C *DWELL 0.30S RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT SETUP RESULTS CALIBRATE CHANGE RUN ERROR Description: The dwell is the time, or distance in vari-speed systems, that the reject device is activated for once the reject device has been reached by the contaminated product.
Page 18
ENABLE REJECT MENUS A to Z L o m a S u p e r s c a n M i c r o I S C ENABLE REJECT? RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT RESULTS SETUP CALIBRATE CHANGE RUN ERROR Description: This command can be used to turn the reject mechanism back on after it has been disabled - see...
Page 19
Note: To gain access to this report, it must first be enabled in the Report Options menu. SAMPLE REPORT --------------------------------------- : LOMA SUPERSCAN MICRO METAL DETECTOR : --------------------------------------- RUN ERRORS REPORT Time : 15-40-44 Date : WED 20 SEP 1995...
Page 20
FALSE THRESHOLD MENUS A to Z L o m a S u p e r s c a n M i c r o I S C *FLS THRSH 2000 RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT RESULTS SETUP CALIBRATE CHANGE RUN ERROR Description:...
Page 21
Note: To gain access to this report, it must first be enabled in the Report Options menu. SAMPLE REPORT --------------------------------------- : LOMA SUPERSCAN MICRO METAL DETECTOR : --------------------------------------- SYSTEM FAULTS REPORT Time : 15-42-14 Date : WED 20 SEP 1995...
Page 22
PV Test is actioned. Test wand sizes can be entered up to 10.00mm this allows entry of all the standard available Loma test wands. An entry must be made to ensure the ferrous test wand is prompted for during a PV Test.
Page 23
FLOW LENGTH MENUS A to Z L o m a S u p e r s c a n M i c r o I S C *FLOW LEN 250MM RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT SETUP RESULTS CALIBRATE CHANGE RUN ERROR Description:...
Page 24
If in doubt, call your nearest LOMA Service Centre.
Page 25
Check that the calibrate pack(s) is representative of production packs (eg. for frozen product, the calibrate pack must be frozen). If in doubt, call your nearest LOMA Service Centre. +POWER REDUCING+ Indicates that the unit is reducing the head power to account for the product effect of the product.
Page 26
GOOD PACKS MENUS A to Z L o m a S u p e r s c a n M i c r o I S C 12675 GOOD PACKS RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT RESULTS SETUP CALIBRATE CHANGE RUN ERROR Description:...
Page 27
If set to unlocked the HEAD POWER will be determined automatically during a calibration cycle, whereas if locked the HEAD POWER setting will not be altered by a calibration cycle. This value should be set to UNLOCK, unless a head power setting has been advised by LOMA which must not be altered.
Page 28
This menu position is only accessible if the Manual Setup menu has been enabled in the User Options menu. On earlier LOMA metal detectors this setting was referred to as the control line settings. The table below shows the relationship between the two:...
Page 29
INTERVAL MENUS A to Z L o m a S u p e r s c a n M i c r o I S C *INTERVAL 01.00 RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT SETUP RESULTS CALIBRATE CHANGE RUN ERROR Description: The interval time is a system relative entry which can be specified in hours and minutes.
Page 30
ISC LIMITS MENUS A to Z L o m a S u p e r s c a n M i c r o I S C ISC LIMITS RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT RESULTS SETUP CALIBRATE CHANGE RUN ERROR Description: This menu gives access to the maximum and minimum threshold limits that can be adopted by the...
Page 31
ISC LIMITS continued MENUS A to Z L o m a S u p e r s c a n M i c r o I S C ISC LIMITS RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT SETUP RESULTS CALIBRATE CHANGE RUN ERROR head power is changed, if calibration occurs when there is no product, or if no products are seen...
Page 32
LOCK KEYBOARD MENUS A to Z L o m a S u p e r s c a n M i c r o I S C LOCK KEYBOARD? RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT RESULTS SETUP CALIBRATE CHANGE RUN ERROR Description: Selection of this command causes the keyboard to become locked, so that no unauthorised users can...
Page 33
I S C LOMA MENU RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT SETUP RESULTS CALIBRATE CHANGE RUN ERROR Description: The contents of this menu are usually only used by LOMA service personnel. Press the [ENTER] key to move into this menu.
Page 34
This menu gives access to details about the software fitted to the machine and the machine ID. Typically, when contacting LOMA for technical support we may wish to know the software version that is fitted to your machine, and this can be obtained from within this menu.
Page 35
MANUAL SETUP MENUS A to Z L o m a S u p e r s c a n M i c r o I S C MANUAL SETUP RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT SETUP RESULTS CALIBRATE CHANGE RUN ERROR Description: Press the [ENTER] key to move into the Manual Setup menu, which will allow the calibration...
Page 36
There should be no need to adjust the maximum threshold during normal operation: do not adjust the value unless you are advised to do so by your LOMA Service Engineer. This menu is only accessible if manual setup has been enabled in the User Options menu.
Page 37
There should be no need to modify the minimum threshold during normal operation: do not adjust the value unless you are advised to do so by your LOMA Service Engineer. This menu is only accessible if the Manual Setup menu has been enabled in the User Options menu.
Page 38
NAME OF PRODUCT MENUS A to Z L o m a S u p e r s c a n M i c r o I S C *NAME ABCDE12345 RUN DETECT CANCEL ENTER CALIBRATE SYSTEM FAULT RESULTS SETUP CALIBRATE CHANGE RUN ERROR Description:...