150
Appendices
Select YES to enable the Performance Validation System (PVS) or (PV
PV test
testing). A number of additional menu positions will appear, which require
data input to set-up for PV testing to operate. Select NO to disable PV
testing.
Select YES to enable fault relay operation when a system fault is generated
Fault on PV
by failure to complete a prompted for PV test. When the conveyor belt motor
controls are wired through this relay the conveyor will stop. Select NO to
disable the system fault relay operation only.
Only appears if photo-eye is set to YES in the Configuration menu, ISC is
Reverse mode
set to OFF, Tracker is set to OFF, Cal trigger is set to OFF and PV test is set
to NO.
Select YES to enable the metal detector Reverse Mode option.
This will then cause an extra menu position to appear in the Product setup
menu, which enables the metal detector to be set to operate either in
NORMAL or REVERSE mode for the product selected.
Wherever possible, Loma will have set the entries within this menu to those required by your
application before the dispatch of your system, or during a commissioning visit by a Loma
engineer if applicable.