Note
U, V, and W-band power sensors are not available from
Hewlett-Packard. For these frequencies use the Anritsu ML83A
Power Meter with the MP715-004 (40 to 60 GHz), the MP716A
(50 to 75 GHz), or the MP81B (75 to 110 GHz) power sensors. Since
the Anritsu model ML83A Power Meter is not capable of internally
storing power sensor cal factors, you must manually correct the
data entry. Refer to example 2 for information on manual entry of
correction data.
Setup Synthesizer Parameters
5.
Turn on the synthesizer and press [PRESET).
The following occurs:
The source module's frequency span is displayed on the
n
synthesizer.
The synthesizer's leveling mode is automatically changed
n
internal to "module leveling".
The source module's maximum specified power is set and
n
displayed.
6.
Press FREQUENCY (START) @ @ 0 @ (GHz, m@ @
7.
Press
Access User Flatness Correction Menu
8.
Press POWER CE). Select Fltness Menu.
9.
Select Delete Menu Delete All. This step insures that the
flatness array is empty.
10.
Press (PRIOR). Leave the delete menu and return to the previous
11.
Select Enter Freq (!?J @ 0 @ (GHz), to enter 26.5 GHz as
the first correction frequency. Enter 31, 32.5, and 40 GHz to
complete the list. Notice that the frequencies are arbitrarily
spaced.
Enter Correction Data into Array
12.
Select Mtr Meas
is now under synthesizer control and is performing the sequence
of steps necessary to generate the correction information at each
frequency point.
If an HP-IB error message is displayed verify that the interface
connections are correct. Check the HP-IB address of the power
meter and ensure that it is the same address the synthesizer is
Menu
Measure Corr All. The power meter
Getting Started Advanced 1-41
from