Pulse Period
Function Group
Menu Map
Description
Programming Codes
See Also
Pulse Rate
Function Group
Menu Map
Description
Programming Codes
See Also
(MOD)
4
This softkey lets you set a value for the internal pulse generator's
pulse period. The range of acceptable values is from 2 pus to 65.5
ms. The factory preset value is 2 ms. When this feature is active, its
current value is displayed in the active entry area.
SCPI:
PULSe:TIMing:PERiod <num>[time suffix] or
Analyzer:
NONE
MODULATION
4
This softkey lets you set the internal pulse generators pulse repetition
rate. The repetition rate can vary from 15.26 Hz to 500 kHz. The
factory preset value is 500 Hz. When this feature is active, its current
value is displayed in the active entry area.
SCPI:
PULse:FREQuency <num>[freq suffix] or
Analyzer:
NONE