Pulse Period; Pulse Rate - HP 8360 User Handbook Manual

8360 series synthesized sweepers
Table of Contents

Advertisement

Pulse Period

Function Group
Menu Map
Description
Programming Codes
See Also

Pulse Rate

Function Group
Menu Map
Description
Programming Codes
See Also
(MOD)
4
This softkey lets you set a value for the internal pulse generator's
pulse period. The range of acceptable values is from 2 pus to 65.5
ms. The factory preset value is 2 ms. When this feature is active, its
current value is displayed in the active entry area.
SCPI:
PULSe:TIMing:PERiod <num>[time suffix] or
Analyzer:
NONE
MODULATION
4
This softkey lets you set the internal pulse generators pulse repetition
rate. The repetition rate can vary from 15.26 Hz to 500 kHz. The
factory preset value is 500 Hz. When this feature is active, its current
value is displayed in the active entry area.
SCPI:
PULse:FREQuency <num>[freq suffix] or
Analyzer:
NONE
HP 8360

Advertisement

Table of Contents
loading

This manual is also suitable for:

83620a83622a83624a83623a83630a83640a ... Show all

Table of Contents