Power Slope - HP 8360 User Handbook Manual

8360 series synthesized sweepers
Table of Contents

Advertisement

Power Offset
Function Group
Menu Map 5
Description
Programming Codes

Power Slope

Function Group
Menu Map 5
Description
Programming Codes
POWER
This softkey changes the mapping of absolute power parameters on
input to the synthesizer. It does not change the RF output produced
by the synthesizer. The equation used to determine the displayed
value is:
Entered or Displayed Power = Hardware Power (ALC) + Active
Offset.
POWer:OFFset:STATe ON]OFF]llO
POWer:OFFset <num>[DB]]MAXimum(MINimum(UP(DOWN
Analyzer:
NONE
[
)
and POWER (FV1ENU).
POWER LEVEL
POWER
This softkey lets you compensate for system, cable, and waveguide
variations due to changes in frequency, by linearly increasing or
decreasing power output as the frequency increases. RF slope values
may range from -2.50 to i-2.50 dB per GHz. The power at the
beginning of the sweep equals the current power level. An asterisk
next to the key label indicates that this feature is active.
POWer:SLOPe:STATe ON(OFFIl]O
POWer:SLOPe <num>[DB](MAXimum]MINimum]UP]DOWN
Analyzer:
functions in the fundamental units of dB/Hz, you program the SL
code SLmdt, where m is 1 (on) or 0 (off); d is the numerical value
HP 8360
User's Handbook

Advertisement

Table of Contents
loading

This manual is also suitable for:

83620a83622a83624a83623a83630a83640a ... Show all

Table of Contents