Program Comments - HP 8360 User Handbook Manual

8360 series synthesized sweepers
Table of Contents

Advertisement

OUTPUT OStimulus;"*IDN?"
210
220
ENTER OStimulus;Id$
230
PRINT Id$
240
PRINT
250 !
260
PRINT "Counter Used
270
OUTPUT OResponse;"*IDN?"
280
ENTER QResponse;Id$
290
PRINT Id$
300
PRINT
310 !
320
OUTPUT OStimulus;":OUTPUT ON"
330 !
340
PRINT
350
PRINT "INPUT [mvI","OUTPUT [kHz]"
360
370
PRINT
380 !
390
FOR Testpoint=First TO Last
400
OUTPUT OStimulus;":SOURCE:VOLT ";VAL$(Testpoint/lOOO);";*OPC?"
410
ENTER OStimulus;Dummy
420
OUTPUT QResponse;":MEAS:FREQ?"
430
ENTER OResponse;Reading
440
PRINT Testpoint,Reading/lOOO
NEXT Testpoint
450
460 !
470
OUTPUT QSource;":OUTPUT OFF"
480
END
Program Comments. Lines 20 to 70: Declare variables and I/O paths
for instruments. I/O paths let you use a name for an instrument in
OUTPUT and ENTER statements,instead of a numeric address.
80 to 100: Assign values to the input test limits in mV.
110 to 130: Clear the instrument HP-IB interfaces.
140 to 160: Reset each instrument to a known measurement state.
170 to 190: Print the test report title.
200 to 310: Query measurement instruments' identifications for test
traceability.
320 to 330: Connect the source output signal to the output
terminals.
340 to 380: Print results table header.
390 to 460: This is the main measurement loop. Line 400 contains
two commands. :SOURce:VOLT sets the output level of the source.
executing. To make an accurate measurement,the source output
must be allowed to settle. When the output has settled, *OPC? places
. . .
Getting Started Programming 1-77