Coupling Factor

Function Group
Menu Map
Description
Programming Codes
See Also
C
CW
Function Group
Menu Map
Description
Programming Codes
See Also
ALC
1
This softkey allows specification of the coupling factor of an external
coupler/detector used to externally level the synthesizer output
power. Negative coupling factor values are required for valid entry.
See "Specifications" for the coupling factor range.
POWer:ALC:CFACtor <num>[dB]~MAXimum~MINimum
SCPI:
Analyzer: NONE
A
ALC
"Externally Leveling the Synthesizer" in Chapter 1.
FREQUENCY
2
This hardkey lets you select a synthesized continuous wave frequency.
When you press Icw), the synthesizer stops sweeping (green SWEEP
LED off) and displays: -->
represents a frequency value. Use either the rotary knob, the step
keys (with or without the left/right arrow keys), or the numerical
entry keys with a terminator key to set the desired value. If a small
change is desired, use the left and right arrow keys to underline the
digit to be changed.
FREQuency[:CW] <num>[freq suffix] or
FREQuency:MODE CW
CW
Analyzer:
"CW Operation and Start/Stop Frequency Sweep" in Chapter 1.
"Programming Typical Measurements" in Chapter 1.
CW:
XXXXX
MHz. Where XXXXX
HP 8360
User's Handbook