HP 8360 User Handbook Manual page 343

8360 series synthesized sweepers
Table of Contents

Advertisement

HP 8360
User's Handbook
FREQ:MODE Affect on Slope
CW or LIST
Rotates around 0 Hz.
Rotates around the start frequency.
STEP
The *RST value is 0.
POWer:SLOPe:STATe ON|OFF|l|O
l
POWer:SLOPe:STATe?
l
Sets and queries the power slope state. *RST value is 0.
POWer:SPAN <num>[DB] lMAXimumlMINimumlUPlDOWN
l
POWer : SPAN? [MAXimum I MINimum]
l
The coupling equations for power sweep are exactly analogous to
those for frequency sweep. Power sweep is allowed to be negative,
unlike frequency sweeps. *RST value is 0.
l
POWer:STARt? [MAXimumlMINimuml
l
Default units and units for query response are determined by the
command UNIT : POWer. The coupling equations for power sweep
are exactly analogous to those for frequency sweep. Power sweep is
allowed to be negative, unlike frequency sweeps.
POWer:STATe ON|OFF|l|O
l
POWer:STATe?
l
Sets and queries the output power on/off state. *RST value is OFF.
l
POWer:STEP:AUTO?
l
Sets and queries the function switch that controls how the power
step size (Power: STEP: INCRement) is determined. if in the automatic
state, then the step size is 1 dB. The *RST setting is ON.
POWer : STEP [ : INCRementI
l
POWer : STEP [ : INCRementI ? [MAXimum I MINimumI
l
Sets and queries the power step size to be used for any node in the
power subsystem that allows UP and DOWN as parameters. Setting this
value explicitly causes POWer : STEP : AUTO OFF.
The *RST setting is 10 dB.
POWer:STOP Qn.un>[lvlsuffix] lMAXimumlMINimumlUPlDOWN
l
POWer : STOP? [MAXimum I MINimumI
l
Sets and queries the ending power for a power sweep. Default units
and units for query response are determined by the command
UNIT:POWer. The coupling equations for power sweep are exactly
Operating and Programming Reference S-45