Power Subsystem
l
Queries the status of any modulation. If any of the modulation states
are on, then it returns a 1, otherwise it returns a 0.
Any place where dBm is accepted as a suffix, any level suffix is
accepted also. In the absence of a suffix, the units are assumed to be
as set by the UNIT :POW command.
POWer:ALC:BANDwidthl:BWIDth <num>[freq
l
POWer:ALC:BANDwidth?|:BWIDth? [MAXimumlMINimuml
l
Sets and queries the ALC bandwidth. This is actually not
continuously variable, so the input is rounded to the nearest possible
switch position. *RST setting is automatically determined since AUT(
is ON.
l
POWer:ALC:BANDwidth|:BWIDth:AUTO?
l
Sets and queries the automatic ALC bandwidth selection switch. The
POWer:ALC:CFACtor <num>CDB1lMAXimumlMINimumlUPlDOWN
l
POWer:ALC:CFACtor? [MINimumlMAXimuml
l
Sets and queries the coupling factor used when the command
POWer : ALC [: SOUFke] is set to DIODe or PMETer.
l
POWer:ALC:SOURce?
l
Sets and queries the ALC leveling source selection switch.
The *RST value is INTernal.
l
POWer : ALC [ : STATe] ?
l
Sets and queries the state switch of the ALC. The positions are :
ON-normal ALC operation
OFF-open loop ALC mode
When on, the power can be programmed in fundamental units as
selected by the UNIT:POWer command.
When off, the power is no longer calibrated in absolute units and is
set in units of dB of arbitrary modulator setting.
POWer:AMPLifier:STATE ONlOFFlOll
l
POWer : AMPLif ier : STATE?
l
Sets and queries the state of the amplifier contained in
the doubler (for those models with a doubler installed).