SWEep:MANual:POINt <num>lMAXimumlMINimum
l
SWEep:MANual:POINt? [MAXimumlMINimum]
l
Sets and queries the step point number to go to and lock. The value
is a unitless value that is limited between 1 and the number of points
requested. This command has no effect on the instrument unless:
the sweep mode is set to manual and the sweep generation is set to
stepped mode. *RST value is 1.
l
SWEep : MANual [ : RELat ive]
l
SWEep : MANual [ : RELat ive] ?
Sets and queries a percent of sweep to go to and lock. This command
has no effect unless: the sweep mode is set to manual and the sweep
generation is set to analog. *RST value is 0.50.
l
SWEep:MARKer:STATe ONlOFFlllO
SWEep:MARKer:STATe?
l
Sets and queries the state of marker sweep. When ON, the frequency
sweep limits are taken to be the positions of marker 1 and marker 2.
SWEep : MARKer : XFER
l
This transfers the values of marker 1 and marker 2 frequencies into
start and stop frequency.
SWEep:MODE AUTOlMANual
l
SWEep : MODE?
l
Selects and queries the manual sweep mode switch.
A U T O
The sweep is under the control of the INIT and
SWEEP
subsystems.
FREQ:MANual, SWEep:MANual[:RELative], and
MANual
SWEep:MANual:POINt control the output.
*RST value is AUTO.
SWEep:POINts <num>lMAXimumlMINimum
l
Sets and queries the number of points in a step sweep. When points
is changed, SWEep : STEP is modified by the equation
STEP = SPAN/POINTS.
Span is normally an independent variable but is changed to STEP
x POINTS if both of these parameters are changed in the same
message. *RST value is 11.
SWEep : STEP Cnum> [f req suffix]
l
SWEep:STEP. 7 [MAXimum I MINimum]
l
Sets and queries the size of each frequency step. : STEP is governed
by the equation
STEP = SPAN/POINTS.