Freq Offset

Function Group
Menu Map
Description
Programming Codes
See Also
HP 8360
User's Handbook
The factory preset value is 1. An asterisk next to the key label
indicates that this feature is active.
FREQuency:MULTiplier <num>IMAXimumJMinimum
FREQuency:MULTiplier:STATe ONlOFFlllO
<num> will be rounded to the nearest integer.
Analyzer:
SHFA <n>
FREQUENCY (MENU), Freq Offset
FREQUENCY
2
This softkey lets you set a frequency offset value and applies it to
all frequency parameters. The frequency offset ranges between and
including fllO.O GHz. Changing the frequency offset value changes
the display but does not affect the output frequency. Frequency
multiplier and offset are related as shown by the following equation:
Entered value or Displayed Frequency = (Frequency
Generated
Multiplier) + Offset value
x
The factory preset value is 0 Hz. An asterisk next to the key label
indicates that this feature is active.
FREQuency:OFFSet <num>IMAXimumIMINimum
FREQuency:OFFSet:STATe ONlOFFlllO
Analyzer:
SHFB <n>[HzlKzlMz(Gzl]
FREQUENCY (MENU), Freq Mult
Operating and Programming Reference F-19