Section 10: Calibrating the Model 9100: Calibration Mode
10.2.2

Mode Selection

10.2.2.1
'Mode' Key
Selection of any one of the Model 9100's five operating 'Modes' is enabled by pressing
the
'Mode'
key at the bottom right of the
in display of the mode selection screen shown below.
Mode Selection
Hz
Select required mode using softkeys
Aux
Mode
TODAYS DATE
PROC
10.2.2.2
'Mode Selection' Screen
The Mode Selection menu screen allows you to select any one of the following operating
modes:-
PROC
Procedure mode
MANUAL
Manual mode
CONFIG
Configure mode
CALIB
Calibration mode
TEST
Selftest mode
Each mode is selected by pressing the appropriate screen softkey, after which a new menu
screen relating specifically to that mode will be displayed. (Note: for the calibration menu
screen to appear, certain other conditions must be satisfied — see opposite.)
10.2.3

Selection of Calibration Mode

The following two conditions must be satisfied before the Calibration mode menu screen
can be accessed:-
a. The rear-panel Calibration Enable switch must be in the 'ENABLE' position.
b. A valid password must have been entered via the front panel keypad.