Fluke 9100 User Handbook Manual page 197

Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

10.3.3.3
Using the Default Calibration Targets
As mentioned earlier, the Model 9100's firmware contains a complete set of recommended
target calibration values, for every hardware configuration of every function that can be
directly calibrated. If you wish to use these 'default' values, rather than the values at which
your 9100 was previously calibrated (assuming that the 'saved' and 'default' values are
different), simply press the DEFLT (Default) softkey which appears in the target
selection screen. The Default Calibration Targets for the selected hardware configuration
will be displayed as shown below. Re-pressing the DEFLT will take you back to the
display of Saved Calibration Targets — i.e. the DEFLT key toggles between the
'Saved' and 'Default' sets of target values.
V
C A L
+ 1 .
V
D E F A U L T C A L I B R A T I O N F A C T O R S
1
— 3 .
2
+ 3 .
TODAYS DATE
V
To select one of the displayed 'default' target values for calibration, press the corresponding
Factor softkey (i.e. the 'Factor 1' key to calibrate at target value one, the 'Factor 2' key
to calibrate at target value two. etc..)
This will transfer you to a Calibrate screen, in which you can determine the output error
and correct it — see Section 10.3.4.
Section 10: Calibrating the Model 9100: Standard Calibration — Basic Sequences
F a c t o r
1
F a c t o r
2
l o w
h i g h
TIME
TARGET
DEFLT
Final Width = 215mm
10.3-7