Fluke 9100 User Handbook Manual page 185

Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

HIGH VOLTAGE
DANGER
sHi
sLo
CJC Pod Module, shown plugged into the D-type socket on the
9100 front panel.
15Vpk
max
(Note that if the 9105 lead set is in use, the calibration can be
!
!
performed with the module plugged into the D-type socket on the
Hi
Lo
end of the connection box underneath the work mat.)
1500Vpk
max
I+
I-
15Vpk
I-
I+
max
Lo
Hi
PRT fully inserted
4-wire connection.
into the Hole in the
end of the Pod
Fig. 10.2.5.1 Setup for Cold Junction Calibration
Section 10: Calibrating the Model 9100: Calibration Mode
Resistance Temperature Detector, or
DMM on 100 /1k Ohms Range.
Use low exitation current (1mA ±20%)
Do not use 2-wire
connection as errors
will be excessive.
10.2-11
Final Width = 215mm