Page 1
Hart Scientific 917X Series Metrology Wells Getting Started ENG 562201 1.888.610.7664 www.calcert.com sales@calcert.com...
Page 2
Limited Warranty & Limitation of Liability Each Metrology Well from Fluke’s Hart Scientific Division (“Hart”) is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year. The warranty period begins on the date of the shipment. Parts, product repairs, and services are warranted for 90 days.
Before You Start Introduction Fluke’s Hart Scientific Division Metrology Wells (9170, 9171, 9172, and 9173) are designed to be stable heat sources that can be used in a laboratory or field environment. With a calibrated display and an optional built-in reference thermometer input (designated with “–R”), Metrology Wells may also be used as a reference standard.
917X Series Metrology Wells Getting Started Symbol Description Fuse PE Ground Hot Surface (Burn Hazard) Read the User’s Guide (Important Information) Canadian Standards Association OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 refers to the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation.
Page 8
Before You Start Safety Information GENERAL DO NOT use this instrument in environments other than those listed in the User’s Guide. Inspect the instrument for damage before each use. DO NOT use the instrument if it appears damaged or operates abnormally. Follow all safety guidelines listed in this guide.
917X Series Metrology Wells Getting Started DO NOT touch the well access surface of the instrument. The block vent may be very hot due to the fan blowing across the heater block of the Metrology Well. The calibration well temperature of the Metrology Well is the same as the actual display temperature, for example, if the instrument is set to 700°C and the display reads 700°C, the well is at 700°C.
CE Comments EMC Directive Fluke’s Hart Scientific Division equipment has been tested to meet the European Electromagnetic Compatibility Directive (EMCEMC Directive, 89/336/EEC). The Declaration of Conformity for your instrument lists the specific standards to which the instrument was tested.
Class B equipment. The instrument was not designed to be used in domestic establishments. Low Voltage Directive (Safety) In order to comply with the European Low Voltage Directive (73/23/EEC), Fluke’s Hart Scientific Division equipment has been designed to meet the IEC 1010-1 (EN 61010-1) and the IEC 1010-2- 010 (EN 61010-2-010) standards.
Page 12
Before You Start Authorized Service Centers Fluke Int’l Corporation Fluke South East Asia Pte Ltd. When contacting a Service Center for support, please have the following information available: • Model Number • Serial Number • Voltage • Complete description of the problem 1.888.610.7664...
Specifications and Environmental Conditions Specifications Specifications and Environmental Conditions Specifications Table 2 Metrology Well specifications Specifications 9170 9171 9172 9173 –45°C to 140°C –30°C to 155°C 35°C to 425°C 50°C to 700°C Range (–49°F to 284°F) (–22°F to 311°F) (95°F to 797°F) (122°F to 1292°F) ±0.1°C full range ±0.1°C: 35°C to 100°C...
917X Series Metrology Wells Getting Started Specifications 9170 9171 9172 9173 Specifications are given with an ambient temperature of 23°C (73.4°F) ± 7°C (13.9°F). Range, display accuracy, axial uniformity, loading effect, cooling time, and heating time are subject to the ambient temperature and may be affected outside the “Full Accuracy” temperature range Refer to the Maintenance section in the User’s Guide regarding maintaining the accuracy of the instrument.
Page 15
Specifications and Environmental Conditions Environmental Conditions • indoor use only 1.888.610.7664 www.calcert.com sales@calcert.com...
Quick Start Unpacking Quick Start Unpacking Unpack the instrument carefully and inspect it for any damage that may have occurred during shipment. If there is shipping damage, notify the carrier immediately. Verify that the following components are present: 9170 • 9170 Metrology Well •...
Sleeves of various sizes are available from Fluke’s Hart Scientific Division. The well must be clear of any foreign object s, dirt and grit before the sleeve is inserted. The sleeve is inserted with the two small tong holes positioned upward.
Quick Start System Setup Turn on the power to the calibrator by toggling the switch on the power entry module. Aft er a brief self-test, the co ntroller should begin normal operation. The main screen will appear within 30 seconds. If the instrument fails to operate, please check the power connection. The display will show the well temperature, and wait for user input before heating or cooling to current set-point.
917X Series Metrology Wells Getting Started limit. Use the numeric keys to set the desired stability limit and press “ENTER” to accept the ne stability limit. Exampl A specific calib ration process requires the instrument be operating within ±0.1°C. “0.1” would be tered into the stability limit parameter.
Quick Start Reference Probe (-R models only) Refer ence Probe (-R models only) The reference probe section of the user manual explains in detail how to set up the reference probe of the calibrator using the front panel keys. The procedure is summarized here. Probe Connection A PRT is the only type of probe that is supported by the reference thermometer in put.
Page 21
If no coefficients can be found, contact the manufacturer or vendor of the probe for assistance. If the probe is out of calibration, Fluke’s Hart Scientific Division offers calibration services. Contact an Authorized Service Center for assistance on obtaining a calibration for the probe.
Parts and Controls Back Panel Parts and Controls This section describes the exterior features of Metrology Wells. Back Panel The following are found on the back of Metrology Wells (see Figure 2). Power Cord Plug (1) The power supply cord attaches to the back panel. Plug the cord into an AC mains supply appropriate for the voltage range as specified in the specifications tables.
917X Series Metrology Wells Getting Started Reference Thermometer Connection (-R models only) (4) The 5-pin DIN connector on the front panel allows a reference probe to be attached to the instrument for use with the reference thermometer function. Switch Connectors (5) The switch hold patented DWF connector posts are located on the right side of the front panel.
Parts and Controls Front Panel Buttons Editing Windows While setting up and operating the instrument, you are often required to enter or select parameters. Editing windows appear on the screen when necessary to show the values of parameters and allow you to change them.
917X Series Metrology Wells Getting Started menu. If a parameter is entered or changed and “EXIT” is pressed before “ENTER”, the change will be canceled. During some operations the “EXIT” key may be used to cancel the operation. If a cutout condition exists, press the “EXIT” key to access the Cutout menu. To reset the cutout, select the “RESET CUTOUT”...
Page 28
Parts and Controls Accessories • 9171-INSE, Insert, E, 9171, metric,0.25 in reference, miscellaneous holes • 9171-INSF, Insert, F, 9171, metric, 0.25 in reference, comparison holes • 91710-INSG, Insert, G, 9171, EA testing • 9171-INSY, Insert, Custom, 9171 • 9171-INSZ, Insert, Blank, 9171 •...