Fluke 9100 User Handbook Manual page 226

Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

10.4.4.6
Calibration of Auxiliary AC Voltage (for Power Option)
Calibration Setup
1. Connections
Ensure that the 9100 is connected to the Standards DMM as shown in Fig. 10.4.3.2, and that both instruments
are powered on and warmed up.
2. 9100
Ensure that the 9100 is in STANDARD CAL mode and then select Power and AC Volts by pressing the front-
panel AUX key followed by the softkey sequence POWER
Refer also to Fig. 10.2.2.
*
This keystroke is only required if one of the oscilloscope calibration options (Option 250 or Option 600) is
fitted.
Table 10.4.4.3: Auxiliary AC Voltage Calibration Points
Hardware Spans
Default
0.000mV to 3.200mV
3.00000mV
3.00000mV
3.00000mV
3.00000mV
3.21mV to 32.00mV
30.0000mV
30.0000mV
30.0000mV
30.0000mV
32.1mV to 320.0mV
300.000mV
300.000mV
300.000mV
300.000mV
0.321V to 3.200V
3.00000V
3.00000V
3.00000V
3.00000V
3.201V to 7.500V
6.00000V
6.00000V
6.00000V
6.00000V
10.4-18
Section 10: Calibrating the Model 9100: AC Voltage Function
*
,
Calibration Targets
Level
Minimum
Maximum
Default
1.80000mV
3.20000mV
40
1.80000mV
3.20000mV
150
1.80000mV
3.20000mV
1000
1.80000mV
3.20000mv
3000
18.0000mV
32.0000mV
40
18.0000mV
32.0000mV
150
18.0000mV
32.0000mV
1000
18.0000mV
32.0000mV
3000
180.000mV
320.000mV
40
180.000mV
320.000mV
150
180.000mV
320.000mV
1000
180.000mV
320.000mV
3000
1.80000V
3.20000
40
1.80000V
3.20000
150
1.80000V
3.20000
1000
1.80000V
3.20000
3000
4.00000V
6.50000V
40
4.00000V
6.50000V
150
4.00000V
6.50000V
1000
4.00000V
6.50000
3000
AUX
W,
, VOLTAGE.
CHANNEL
MODE
Frequency
Minimum
Maximum
35
45
140
160
900
1100
2500
3000
35
45
140
160
900
1100
2500
3000
35
45
140
160
900
1100
2500
3000
35
45
140
160
900
1100
2500
3000
35
45
140
160
900
1100
2500
3000