When you have finished calibrating one target value you will be returned to the function
screen, from which you can:
either calibrate the remaining target values for the selected hardware configuration;
or
select a new hardware configuration by changing the displayed output value.
Before deciding to change from 'Saved' calibration target values back to the 'Default'
target values, consider that there may be good reasons for choosing values other than the
default values at the previous calibration. Your organization's calibration standards may
be uniquely characterized at particular target values (for example, three series connected
standard cells which provide a calibration point voltage of 3.05426V rather than the
default value of the 3.00000V for the 0.32001V to 3.20000V DC Voltage hardware
configuration). Changing the default target values back to 3.00000V may reduce the
usefulness of historical calibration records for this 9100 unit, if it normally uses
3.05426V as the target value.