6.7.5
Other Commands
These are existing 9100 commands and are shown here as they are used to manipulate parameters associated with the power option.
6.7.5.1
[SOURce]:FREQuency[:CW|FIXed(?)
This command sets the frequency of the currently selected waveform. The units of this command are assumed to be Hertz. The CW
and FIXed optional parameters are included to keep with the SCPI definition of the frequency command:
"The Continuous Wave or FIXed node is used to select a frequency of a non-swept signal".
• This command is valid in AC waveshapes only.
• In Harmonics, it sets the fundamental frequency.
6.7.5.2
[SOURce]:PHASe[:ADJust(?)
This command is used to set the phase angle of the 9100 output, with respect to 'Phase Lock In', when operation in either AC
Voltage, Current or Power functions.
• The range of the <dnpd> is +180.0 to -180.0. Values outside this will generate a Data out of range error.
6.7.5.3
[SOURce]:PHAse:INPut[:STATe](?)
The 9100 AC Voltage or AC Current output can be phase-shifted from a reference-phase input via the 'PHASE LOCK IN' plug
on the rear panel (by a phase angle set by the PHASe:ADJust command).
6.7.5.4
[SOURce]:OUTPut:STATe]
The 9100 AC Voltage or AC Current output phase used as a reference-phase output via the 'PHASE LOCK OUT' plug on the
rear panel. This command switches the reference phase on or off.
6-62
Section 6: 9100 System Operation — SCPI Language
<dnpd>
<dnpd>
<cpd>{ON|OFF|0|1}
<cpd>{ON|OFF|0|1}