10.4.10.3
Calibration Procedure
1. Select the required hardware configuration by setting the 9100 output to a suitable
value that uses the hardware configuration.
2. Set the continuity tester to the appropriate measurement range
3. Press the TARGET softkey to display the hardware configuration's target
selection screen.
4. Press the FACTOR softkey corresponding to the required target value to display the
calibration screen for this value.
5. If the displayed target value is not the same as the value of your standard resistor for
this calibration point, press the TAB key to position the cursor on the target value and
edit this value to that of your standard resistor.
6. Press the TRANSFER softkey to transfer the target calibration point to the 9100
output control display.
7. Connect the continuity tester to the standard resistor, press the TEST button and note
its reading (allow for the settling time of the test meter).
8. Release the TEST button and reconnect the continuity tester to the 9100 using the
same test leads.
9. Press the ON key to turn the 9100 output on. Press the TEST button on the continuity
tester and note the reading.
10. Adjust the 9100 output amplitude until the reading on the continuity tester is the same
as that noted in step (7)
11. When you are satisfied with the measurement, press the CALIB softkey to generate
and implement the correction factor required by the 9100.
12. Press the OFF key to turn the 9100 output off.
13. Repeat steps (2) through (12) for both of the target values displayed in the target
selection screen.
14. Repeat steps (1) through (13) for each of the hardware configurations detailed in the
table opposite.
Section 10: Calibrating the Model 9100: Continuity Function