Final Width = 215mm
10.4-58
Section 10: Calibrating the Model 9100: Continuity Function
10.4.10.2
Calibration Equipment Required
A set of traceably calibrated standard resistors at values between the minimum and
maximum targets defined in the following table, capable of passing at least 350mA
DC.
1
A 3
/
digit (or better) continuity test meter (for example a Yokogawa 2426). This is
2
used to make transfer measurements between the standard resistors and the
9100.