10.4.9.4
Calibration Setup
1. Connections
Ensure that the 9100 is connected to the continuity tester as shown in Fig. 10.4.10.1 and that both instruments
are powered on and warmed up.
2. 9100
Ensure that the 9100 is in STANDARD CAL mode and then select the Continuity function by pressing the
AUX key on the right of the front panel followed by the
Table 10.4.9.2: Continuity Hardware Configurations and Calibration Points
Output span covered by
Suitable output Recommended Minimum Maximum Factor
hardware
to select h/w
configuration
configuration
00.0000
40.0000
10
40.001
400.000
100
0.40001k
4.00000k
1k
10.4-60
Section 10: Calibrating the Model 9100: Continuity Function
Number
1.00
0.01
7.50
1
30.00
10.00
39.0
2
60.00
41.00
75.00
1
300.0
100.0
390
2
600.0
410.0
750.0
1
3.000k
1.000k
3.900k
2
softkey.