Fluke 9100 User Handbook Manual page 248

Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

Final Width = 215mm
10.4.7.6
Calibration Setup
1. Connections
Ensure that the 9100 is connected to the Standards DMM as shown in Fig. 10.4.7.1, and that both instruments
are powered on and warmed up.
2. 9100
Select the Resistance Function by pressing the ' ' key on the right of the front panel.
10.4.7.7
Calibration of 'UUTi High' Source Current Hardware Configurations
If both UUTi High and UUTi Low source current hardware configurations are being calibrated, the UUTi High source current
calibrations shown in Table 10.4.7.2 below should be performed before the UUTi Low source current calibrations.
Table 10.4.7.2: Resistance — 'UUTi High' Source Current Hardware Configurations and Calibration Points
Output span
Output
covered by
to select
hardware
hardware
configuration
config.
Default
00.0000
10
1.0000
to 40.0000
30.0000
40.001
100
60.000
to 400.000
300.000
0.40001k
1k
0.60000k
to 4.00000k
3.00000k
4.0001k
10k
6.0000k
to 40.0000k
30.0000k
40.001k
100k
60.000k
to 400.000k
300.000k
0.40000M
1M
0.60000M
to 4.00000M
3.00000M
4.0001M
10M
6.0000M
to 40.0000M
30.0000M
40.001M
100M
60.000M
to 400.000M
300.000M
Notes:
[1] This calibration point is limited by the maximum source current. If the actual source current is less than the indicated maximum, the output
value can be raised in inverse proportion. For example, a 50nA source current would allow the calibration point to be increased to 160M .
10.4-40
Section 10: Calibrating the Model 9100: Resistance Function
Calibration Targets
Amplitude
Minimum
Maximum
Recommended
0.0100
7.5000
0.1000
10.0000
39.0000
38.0000
41.000
75.000
45.000
100.000
390.000
380.000
0.41000k
0.75000k
0.45000k
1.00000k
3.90000k
1.95000k
4.1000k
7.5000k
4.5000k
10.0000k
39.0000k
38.0000k
41.000k
75.000k
45.000k
100.000k
390.000k
380.000k
0.41000M
0.75000M
0.45000M
1.00000M
3.90000M
3.80000M
4.1000M
7.5000M
4.5000M
10.0000M
39.0000M
38.0000M
41.000M
70.000M
45.000M
76.500M
390.000M
80.000M
Factor
Number
Source Current
Minimum Maximum
500 A
2.0mA
1
500 A
2.0mA
2
500 A
2.0mA
1
500 A
2.0mA
2
500 A
2.0mA
1
500 A
2.0mA
2
50 A
200 A
1
50 A
200 A
2
5.0 A
20 A
1
5.0 A
20 A
2
500nA
2.0 A
1
500nA
2.0 A
2
50nA
200nA
1
50nA
200nA
2
50nA
100nA
1
50nA
100nA
2
[1]