Section 10: Calibrating the Model 9100: Standard Calibration — Basic Sequences
At this point you have two options in selecting the target calibration points for use in
calibrating this hardware configuration.
You can:-
1) use the same target points (the Saved Calibration Targets) at which this hardware
configuration was last calibrated — see Section 10.3.3.2.
2) use default target points which are pre-programmed into the Model 9100's firmware
— see Section 10.3.3.3.
Note that the 9100 also provides you with a third option, which is to change the existing
calibration targets to new values. However, this option only becomes available when you
move on to the Calibration screen — see Section 10.3.4.
At any time you can return to the introductory function screen by re-pressing the
TARGET softkey.
10.3.3.2
Using the Saved Calibration Targets
To use the 'Saved Calibration Targets' displayed on the target selection screen, there is
no need to modify that screen. All that is required, to select one of the displayed target
values, is to press the corresponding Factor softkey (i.e. the 'Factor 1' key to calibrate
at target value one, the 'Factor 2' key to calibrate at target value two. etc..)
This will transfer you to a 'Calibrate' screen, which you can use to determine the output
error, and correct it — see Section 10.3.4.
When you have finished calibrating one target value you will be returned to the function
screen, from which you can:
either calibrate the remaining target values for the selected hardware configuration;
or
select a new hardware configuration by changing the displayed output value.