Fluke 9100 User Handbook Manual page 262

Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

Final Width = 215mm
10.4-54
Section 10: Calibrating the Model 9100: Insulation Resistance Function
10.4.9.4
Calibration Setup
1. Connections
Ensure that the 9100 is connected to the insulation tester as shown
in Fig. 10.4.9.1 and that both instruments are powered on and
warmed up.
2. 9100
Ensure that the 9100 is in STANDARD CAL mode and then select
the Insulation Resistance function by pressing the AUX key on the
right of the front panel followed by the
Table 10.4.9.1: Insulation Resistance Hardware Configurations and Calibration Points
Output span covered by
Suitable output Recommended Minimum Maximum Factor
hardware
to select h/w
configuration
configuration
100.0K
299.9K
200.0K
300.0K
2.999M
2.000M
3.000M
29.99M
20.00M
30.00M
299.9M
200.0M
300.0M
2.000G
0.600G
softkey.
Number
110.0K
100.0K
120.0K
1
270.0K
240.0K
299.9K
2
330.0K
300.0K
360.0K
1
1.000M
0.900M
1.100M
2
3.300M
3.000M
3.600M
1
10.00M
9.000M
11.00M
2
33.00M
30.00M
36.00M
1
100.0M
90.00M
110.0M
2
660.0M
600.0M
720.0M
1
1.800G
1.600G
2.000G
2