Fluke 9100 User Handbook Manual page 107

Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

7.6.5
AC Current Accuracy (Square-Wave)
Frequency
Output Current Span
[2]
Band
RMS
(Hz)
10 - 1k
0 A - 45.08 A
10 - 1k
45.08 A - 4.508mA
10 - 1k
4.508mA - 45.08mA 4.525mA - 45.25mA
10 - 1k
45.08mA - 450.8mA 45.25mA - 452.5mA
10 - 100
0.4508A - 3.200A
10 - 100
3.200A - 18.00A*
10 - 65
4.508A - 32.00A
10 - 65
32.00A - 180.0A*
10 - 65
22.54A - 160.0A
10 - 65
160.0A - 900.0A*
= With output 'ON', maximum duty cycle of (>0.528FS : 0.528FS) is (1 : 4).
Continuous output >0.528FS will automatically reduce to <0.528FS after 2 Minutes.
† = Total uncertainty includes compliance errors for Voltage 0.5VRMS.
Above 0.5V, add appropriate compliance error, except for Outputs marked
= Refers to accuracy at 9100 output terminals. With Option 200 10 turn coil connected, then at the output from the coil, add 0.2% of output from
coil for uncertainty of coil.
= Refers to accuracy at 9100 output terminals. With Option 200 50 turn coil connected, then at the output from the coil, add 0.2% of output from
coil for uncertainty of coil.
= For frequencies <40Hz, compliance voltage is reduced by 0.5V RMS.
7.6.6
AC Current Accuracy (Impulse-Wave)
Frequency
Output Current Span
[2]
Band
RMS
(Hz)
10 - 1k
0 A - 23.79 A
10 - 1k
23.79 A - 2.379mA
10 - 1k
2.379mA - 23.79mA 4.525mA - 45.25mA
10 - 1k
23.79mA - 237.9mA 45.25mA - 452.5mA
10 - 100
0.2379A - 2.379A
10 - 100
2.379A - 15.00A*
10 - 65
2.379A - 23.79A
10 - 65
23.79A - 150.0A*
10 - 65
11.90A - 118.9A
10 - 65
118.9A - 750.0A*
= With output 'ON', maximum duty cycle of (>0.7FS : 0.7FS) is (1 : 4).
Continuous output >0.525FS will automatically reduce to <0.525FS after 2 Minutes.
† = Total uncertainty includes compliance errors for Voltage 0.5VRMS.
Above 0.5V, add appropriate compliance error, except for Outputs marked
= Refers to accuracy at 9100 output terminals. With Option 200 10 turn coil connected, then at the output from the coil, add 0.2% of output from
coil for uncertainty of coil.
= Refers to accuracy at 9100 output terminals. With Option 200 50 turn coil connected, then at the output from the coil, add 0.2% of output from
coil for uncertainty of coil.
= For frequencies <40Hz, compliance voltage is reduced by 0.5V RMS.
Accuracy †
Peak
(% Output + Floor)
1 Year — Tcal
[1]
5 C
0V - 45.25 A
0.21 + 1.8 A
45.25 A - 4.525mA
0.21 + 0.6 A
0.21 + 6.4 A
0.24 + 64 A
0.4525A - 3.212A
0.30 + 960 A
3.212A - 18.07A
0.4 + 13.8mA
4.525A - 32.12A
1.0 + 16.8mA
32.12A - 180.7A
1.2 + 162mA
22.63A - 160.6A
1.0 + 84mA
160.6A - 903.5A
1.2 + 0.82A
and
Accuracy †
Peak
(% Output + Floor)
1 Year — Tcal
[1]
5 C
0V - 45.25 A
0.42 + 2.7 A
45.25 A - 4.525mA
0.42 + 0.9 A
0.42 + 9.6 A
0.48 + 96 A
0.4525A - 4.525A
0.60 + 1.44mA
4.525A - 28.53A
0.80 + 20.7mA
4.525A - 45.25A
0.80 + 25.2mA
45.25A - 285.3A
1.20 + 243mA
22.63A - 226.3A
0.80 + 126mA
226.3A - 1426A
1.20 + 1.23A
and
Ratios and Factors based on
Voltage
Square-Wave Unit-Value Peak
Compliance
(VRMS)
Peak:
4.0
Peak to Peak:
4.0
RMS:
4.0
4.0
Mean:
2.2
Crest Factor:
2.2
Form Factor:
2.2
2.2
2.2
2.2
.
Ratios and Factors based on
Voltage
Impulse-Wave Unit-Value Peak
Compliance
(VRMS)
Peak:
3.0
Peak to Peak:
3.0
RMS:
3.0
3.0
Mean:
1.8
Crest Factor:
1.8
Form Factor:
1.8
1.8
1.8
1.8
.
Section 7: Model 9100: Specifications
1.0000
2.0000
0.9962
0.9958
1.0038
1.0004
1.0000
2.0000
0.5270
0.3333
1.8974
1.5811
7-11
Final Width = 215mm