Fluke 9100 User Handbook Manual page 216

Universal calibration system
Hide thumbs Also See for 9100:
Table of Contents

Advertisement

Final Width = 215mm
10.4.3.4
Calibration Setup
1. Connections
Ensure that the 9100 is connected to the Standards DMM as shown in Fig. 10.4.3.1, and that both instruments
are powered on and warmed up.
2. 9100
Ensure that the 9100 is in STANDARD CAL mode and then select the DC Voltage function by pressing the
'
' key on the right of the front panel.
V
Table 10.4.3.1: DC Voltage (Zero to 320V) Hardware Configurations and Calibration Points
Output span covered
Suitable output to
by hardware configuration
select hardware
configuration
000.000mV to 320.000mV
0.32001V to 3.20000V
3.2001V to 32.0000V
32.001V to 320.000V
Table 10.4.3.2: DC Voltage (320V to 1050V) Hardware Configurations and Calibration Points
Output span covered
Suitable output to
by hardware configuration
select hardware
configuration
320.01V to 1050.00V
10.4-8
Section 10: Calibrating the Model 9100: DC Voltage Function
Calibration Targets
Default
100mV
000.000mV
+300.000mV
1V
-3.00000V
+3.00000V
10V
-30.0000V
+30.0000V
100V
-300.000V
+300.000V
Calibration Targets
Default
1000V
-1000.00V
+1000.00V
Minimum
Maximum
-0.080mV
10.000mV
180.000mV
320.000mV
-3.20000V
-1.80000V
+1.80000V
+3.20000V
-32.0000V
-18.0000V
+18.0000V
+32.0000V
-320.000V
-180.000V
+180.000V
+320.000V
Minimum
Maximum
-1050.00V
-700.00V
+700.00V
+1050.00V
Factor
Number
1
2
1
2
1
2
1
2
Factor
Number
1
2