10.4.8.4
Calibration Setup
1. Connections
Ensure that the 9100 is connected to the capacitance meter as shown in Fig. 10.4.8.1, and that both instruments
are powered on and warmed up.
2. 9100
Ensure that the 9100 is in STANDARD CAL mode and then select the Capacitance function by pressing the
'
Aux
' key on the right of the front panel followed by the '
10.4.8.5
Calibration of 'UUTi Low' Source Current Hardware Configurations
For calibration of UUTi Low source current hardware configurations use Table 10.4.8.1.
Table 10.4.8.1: Capacitance UUTi Low Hardware Configurations and Calibration Points
Output span covered
Suitable output to
by hardware configuration
select hardware
configuration
0.5000nF to 4.00000nF
4.0001nF to 40.0000nF
40.001nF to 400.000nF
0.40001 F to 4.00000 F
4.0001 F to 40.0000 F
40.001 F to 400.000 F
0.40001mF to 4.00000mF
4.0001mF to 40.0000mF
Calibration Targets
Recommended
1nF
0.60000nF
3.00000nF
10nF
6.0000nF
30.0000nF
100nF
60.000nF
300.000nF
1 F
0.60000 F
3.00000 F
10 F
6.0000 F
30.0000 F
100 F
60.000 F
300.000 F
1mF
0.60000mF
3.00000mF
10mF
6.0000mF
30.0000mF
Section 10: Calibrating the Model 9100: Capacitance Function
' screen key adjacent to the display.
Minimum
Maximum
0.52000nF
0.80000nF
2.50000nF
3.90000nF
4.1000nF
8.0000nF
25.0000nF
39.0000nF
41.000nF
80.000nF
250.000nF
390.000nF
0.41000 F
0.80000 F
2.50000 F
3.90000 F
4.1000 F
8.0000 F
25.0000nF
39.0000 F
41.000 F
80.000 F
250.000 F
390.000 F
0.41000mF
0.80000mF
2.50000mF
3.90000mF
4.1000mF
8.0000mF
25.0000mF
39.0000mF
Factor
Number
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
10.4-47
Final Width = 215mm