Section 10: Calibrating the Model 9100: Standard Calibration — Basic Sequences
When you select one of these calibration points by pressing its corresponding Factor key,
the target calibration point voltage and frequency will be displayed below the current
output voltage and frequency settings as shown in the following screen illustration.
V
C A L
.
.
3 .
V
.
TODAYS DATE
V
TRANSFR
You should now set the current output voltage and frequency settings to the target values
by pressing the TRANSFR softkey, as you would when calibrating any other function.
If you now wish to change the calibration point frequency, this must be done by directly
editing the output frequency setting, not by editing the displayed target frequency.
To do this you must first transfer cursor control to the displayed output frequency setting
by pressing the Tab
key. (Note: depending on the current position of the digit cursor
you may need to press the Tab
key one or two times to shift it to the output frequency
setting.) You can now alter the frequency setting using the same digit edit (cursor
controls/spinwheel) method or direct edit (numeric keypad) method used to change
frequency values in normal operation of the Model 9100 — see Section 3.4 of the Model
9100 Universal Calibration System User's Handbook — Volume 1 — Operation.
x
V
kHz
Z E R O
k H z
TIME
TARGET
CALIB