2. Disconnect the Ribbon Cable
a. Ensure that 9100 Power
is still OFF
b. Locate the ribbon cable connector and latch levers on the
Interconnection PCB attached to the rear panel:
i. Carefully push both latch levers outwards at the same
time. This will ease the connector out of its housing.
ii. Release the connector and pull the ribbon cable
towards the front of the instrument to reveal the fuse
holders beneath.
3. Fuse Removal
a. Ensure that 9100 Power
is still OFF
b. Locate the three bayonet fuse caps (Fig. 8.2.3):
i. Release the relevant fuse cap(s) by pushing down,
turning counter-clockwise.
ii. Remove the fuse(s) by pulling carefully.
4. Fuse Testing
a. Check each removed fuse for continuity and correct type.
b. Decide whether the fuse is to be re-fitted or replaced.
5. Fuse Fitting
WARNING
MAKE SURE THAT ONLY FUSES WITH THE REQUIRED
RATED CURRENT AND OF THE SPECIFIED TYPE ARE
USED FOR REPLACEMENT.
THE FUSE-HOLDERS ARE NOT TO BE SHORT-
CIRCUITED. THIS PRACTICE WILL RENDER THE
WARRANTY VOID.
a. Ensure that 9100 Power
is still OFF
b. Locate the three bayonet fuse holders (Fig. 8.2.3):
i. Fit each correct fuse by inserting it into the holder,
and carefully push the fuse home.
ii. Secure the relevant fuse cap(s) by pushing down and
turning clockwise.
6. Reconnect the Ribbon Cable
.
a. Ensure that 9100 Power
b. Locate the ribbon cable connector across the top of the
instrument, and into its housing on the Interconnection
PCB attached to the rear panel:
i. Carefully push the connector into its housing. This
will cause the latch levers to move together and latch
as the connector is driven fully home.
ii. Finally, push the latch levers together to ensure that
both latches are fully home, with the connector
correctly housed.
.
7. Refitting the Top Cover
Carefully refit the top cover into the slots in the side
extrusions, with the front edge immmediately behind the
front bezel; push forward to locate inside the bezel, and
secure to the rear panel using the two screws. Refit and
secure the corner blocks.
.